-
Federico Vaga authored
We have always been useing an uncalibrated DAC value whenever we were using 'test data' or the input configuration was in open-drain. We recently discovered that this is not true, and actually the voltage range on the input channel is 1V. For this reason the DAC value still needs to be fixed with eeprom calibration data. Signed-off-by: Federico Vaga <federico.vaga@cern.ch> Acked-by: Dimitris Lampridis <Dimitris.Lampridis@cern.ch>
5f5454af